                   Important
		   =========
If you do use a basis set with 'all' in its name, be sure to use NOMIT &
CHOMIT (Section 9.8.1) to omit all but those expected to have significant
signals in the spectrum being analyzed.


                    Contents
		    ========
(Metabolite names are in Table 8.1 of ~/.lcmodel/doc/manual.pdf)

The recommended files (i.e., without 'all' in their names) have Ala, Asp, Cr,
PCr, GABA, Glc, Gln, Glu, GPC, PCh, GSH, Ins (or mI), Lac, NAA, NAAG, Scyllo &
Tau.
Those for TE>100ms have Asp, GABA & Glc omitted.

Those with 'all' in their names have all of the above metabolites plus 
Acn, Act, Asc, Cit, Eth, Glyc, 2HG (2-hydroxyglutarate), PAl, PE, Pgc, Pyr, 
Ser, Suc, Thr & Val.

You can also use Section 11.7 to simulate further singlets.


                  Compatibility
		  =============
The TE=0 basis set might be a fair approximation for STEAM (or PRESS) with
TE<11ms.

They would be an excellent approximation if their field strength is within
5MHz of yours (e.g., for 7T, 298MHz and 300MHz are both fine).  LCModel